CryoChamber
Semetrol designed cryochambers were designed by a user with years of experience, considering ease of sample mounting and contact and other factors. Speed of response, thermal gradients, vibration, and vacuum stability are all important factors to consider in the cryochamber design.
Temperature controlled chambers are designed and tested to quickly obtain stable operation. Many of the measurements require stabilization at 1-5 Kelvin steps over several hundred Kelvin. Optimizing the speed of response is a high priority to obtain timely results. Stabilization is done using a PID controller. The user is able to modify the PID values as necessary.
Semetrol’s stage design places the heater and temperature sensor close together, and also close to the sample. High thermal conductivity grade copper is used for the stage. Temperature testing not only optimizes the response time to a step increase or decrease in temperature, but also tests the time it takes for the sample to get to a stable temperature near the set point.
Samples with low thermal conductivity, or with a thermal impedance between the sample active region and the stage, such as with solar cells on glass substrates, may take more than a minute to stabilize for a 5K step. Other samples such as SiC stabilize more quickly. The temperature lag can be characterized using your samples. Thermal radiation shielding is also used to minimize the significant effect of the temperature difference between the sample and chamber walls.
Each chamber is tested for pump down rate and leak-up rate. The data is supplied to the end user to compare the performance after years of use. A slow pump down rate may mean contamination, e.g. from handling the sample with fingers or from high humidity. A leak-up rate exceeding the initial, new condition, may indicate an O-ring that is contaminated with dust, for example. Having the data available allows you to quickly answer questions such as “Is the pump working as it should?”, “Am I pumping on the chamber long enough before making measurements?”.
The vacuum gauge readout has a computer link and program to graph the vacuum level, and rate of change. The measurements are very sensitive. Leaks in lab vacuum equipment may be easily found using the gauge and program, e.g. by coating possible leak points with a volatile liquid and watching for a change in pressure.
The chamber has four ports: temperature control feedthrough, 4 BNC connectors, vent valve, and vacuum port. Other configurations are possible. Components are selected for hermeticity and reliability.
Each cryochamber is tested for vacuum leak rates, thermal response over the full temperature range, and vibration levels. All test results are available to the end user to compare behavior after years of use.